Linux 6.7-rc7
[linux-modified.git] / Documentation / ABI / testing / sysfs-bus-iio-cdc-ad7746
1 What:           /sys/.../iio:deviceX/in_capacitableY_calibbias_calibration
2 What:           /sys/.../iio:deviceX/in_capacitableY_calibscale_calibration
3 KernelVersion:  6.1
4 Contact:        linux-iio@vger.kernel.org
5 Description:
6                 Write 1 to trigger a calibration of the calibbias or
7                 calibscale. For calibscale, a full scale capacitance should
8                 be connected to the capacitance input and a
9                 calibscale_calibration then started.  For calibbias see
10                 the device datasheet section on "capacitive system offset
11                 calibration".